Transmission Measurement
This measurement will yield the transmission characteristics of a two-port network. The built-in tracking
generator serves as a signal source. It is connected to the input connector of the DUT. The input of the
R&S ESCI is fed from the output of the DUT.
0
12
3
45
6
789
.
-
ESC
1142.8007.03
FCTN
ENTER
CANCEL
GEN OUTPUT 50
AF O UTPUT
MEAS T RIG
FREQ
MKR
AMPT
SPAN
MKR
MKR
BW SWEEP
TRACE
LINES
DISP
FILE
GHz
MHz
kHz
Hz
-dBm
dBm
dB
dB..
CAL
SETUP
PRESET
HCOPY
TEST RECEIVER 9kHz . . . 3GHz
..
ESPI
BACK
s
V
m
s
mV
µ
s
µ
V
n
s
nV
PREV
RF INP UT
50
EXT MIXER
MA
X
+30 dBm / 50V DC
NEXT
KEYBOARDPROBEPOWER
MAX 0V DC
MADE I N GERMANY
.
RF
INPUT
GEN
OUTPUT
DUT
Fig. 4-25 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (eg frequency response
of connecting cables).
Calibration of Transmission Measurement
NETWORK menu:
REF VALUE
POSITION
REF VALUE
SOURCE
CAL
NORMALIZE
CAL
TRANS
CAL REFL
OPEN
RECALL
CAL REFL
SHORT
The SOURCE CAL softkey opens a submenu comprising
of the calibration functions for the transmission and
reflection measurement.
The calibration of the reflection measurement (CAL
REFL...) and its mechanisms are described in separate
sections.
To carry out a calibration for transmission measurements
the whole test setup is through-connected (THRU).