R&S ESCI Configuration/Setup
1166.6004.12 4.209 E-3
Activating Transducer Factors and Transducer Sets
The TRANSDUCER softkey opens a submenu in which already defined transducer factors or sets can
be activated or deactivated, new transducer factors or sets can be generated or existing transducer
factors or sets can be edited. Tables with the defined transducer factors and sets are displayed. The
table (factor or set) in which a transducer is active is set.
By switching on the transducers all the level settings and outputs are automatically made in the unit of
the transducer. A change of the unit in menu AMPT is no longer possible as R&S ESCI together with the
transducer used is regarded as a measuring instrument. Only if the transducer has the unit dB, can the
original unit at R&S ESCI be maintained and changed.
If a transducer factor is active, "TDF" is displayed in the column of the enhancement labels and "TDS" if
the transducer set is active.
After switching off all the transducers, R&S ESCI continues to use the unit which was selected before a
transducer was switched on.
In the analyzer mode, an active transducer for a sweep is uniquely calculated for each point displayed
after its setting and added to the result of the level measurement during the sweep. If the sweep range
is changed, the correction values are recalculated. If several measured values are combined, only a
single value is taken into consideration.
In the receiver mode, the transducer is also calculated for a set scan. The transducer is uniquely
calculated for each frequency point and added to the result of the level measurement as the
measurement results are stored internally and can be zoomed subsequently.
If an active transducer factor/set is not defined over the whole sweep or scan range during the
measurement, the missing values are replaced by zero.
SETUP menu:
TRANSDUCER
TRANSDUCER
FACTOR
TRANSDUCER
SET
NEW
EDIT
DELETE
VIEW
TRANSDUCER
PAGE
UP
PAGE
DOWN
The TRANSDUCER softkey opens a submenu for editing
existing transducer factor and sets or creating new
transduser factor and sets.
Tables show the settings of already existing factors and
sets as well as the settings of the active factor and set.