R&S ESCI Tracking Generator - Option R&S FSP-B9
1166.6004.12 4.273 E-1
Tracking Generator Settings
The NETWORK hotkey opens a menu for selecting the functions of the tracking generator.
SOURCE
ON OFF
SOURCE
POWER
SOURCE
CAL
FREQUENCY
OFFSET
POWER
OFFSET
NETWORK
CAL
TRANS
EXT AM
EXT FM
EXT I/Q
CAL REFL
SHORT
CAL REFL
OPEN
REF VALUE
POSITION
REF VALUE
NORMALIZE
RECALL
MODULATION
OFF
MODULATION
Note: Additional softkeys are available in the displayed menus for controlling an external
generator if option External Generator Control FSP-B10 is fitted. For detailed information
see section 'External Generator Control Option FSP-B10'.
SOURCE
ON OFF
The SOURCE ON/OFF softkey switches the tracking generator on or off.
Default setting is OFF
Note:
• When the tracking generator is switched on the maximum stop frequency
is limited to 3 GHz. This upper limit is automatically reduced by a
frequency offset set up for the tracking generator.
• In order to meet the data sheet accuracy for measurements with active
tracing generetor the start frequency must be set to
≥
3 x Resolution
Bandwidth.
• The minimum sweeptime for measurements with data sheet accuracy is
100 ms in frequency domain (span > 0 Hz). Selecting a sweeptime
below this limit will result in the sweeptime indicator field SWT being
supplied with a red asterisc and the message UNCAL being displayed.
• FFT filters (FILTER TYPE FFT in BW menu) are not available when the
tracking generator is active.
IEC/IEEE-bus command: OUTP:STAT ON