R&S ESCI Configuration/Setup
1166.6004.12 4.235 E-3
General Service Functions
SETUP SERVICE submenu:
INPUT
RF
The INPUT RF softkey switches the input of the R&S ESCI to the input
connector (normal position).
After PRESET, RECALL or R&S ESCI power on, the INPUT RF is always
selected.
IEC/IEEE-bus command: DIAG:SERV:INP RF
INPUT
CAL
The INPUT CAL softkey switches the RF input of the R&S ESCI to the
internal calibration source (128 MHz) and activates the data entry of the
output level of the calibration source. Possible values are 0 dB and –30 dB.
IEC/IEEE-bus command: DIAG:SERV:INP CAL;
DIAG:SERV:INP:CSO 0 DBM
ENTER
PASSWORD
The ENTER PASSWORD softkey allows the entry of a password.
The R&S ESCI contains a variety of service functions which, if incorrectly
used, can affect correct operation of the receiver. These functions are
normally not accessible and are only usable after the entry of a password
(see instrument service manual).
IEC/IEEE-bus command: SYST:PASS "Password"
SETUP SERVICE NEXT submenu:
CAL GEN
128 MHZ
The CAL GEN 128 MHZ softkey selects a sinusoidal signal at 128 MHz as
output signal for the internal calibration source. The internal pulse generator
will be switched off.
CAL GEN 128 MHZ is the default setting of the R&S ESCI.
IEC/IEEE-bus command: DIAG:SERV:INP:PULS OFF
PULSE
xx XXX
The PULSE xx softkey switches the selected pulse generator on.
IEC/IEEE-bus command: DIAG:SERV:INP CAL
DIAG:SERV:INP:PULS ON;
DIAG:SERV:INP:PRAT 2 MHZ