Configuration/Setup R&S ESCI
1166.6004.12 4.222 E-3
User Port Configuration
The instrument provides a parallel interface, which is 8 bits wide. Over these ports, arbitrary bit patterns
can be output or input.
CONFIGURATION SETUP-GENERAL SETUP submenu:
USER PORT
USER PORT
IN OUT
PORT 0
0 1
PORT 1
0 1
PORT 2
0 1
PORT 3
0 1
PORT 4
0 1
PORT 5
0 1
PORT 6
0 1
PORT 7
0 1
The USER PORT softkey opens the sub menu for setting the
direction of the data transmission and for entering the values
of the individual ports.
USER PORT
IN OUT
The USER PORT IN/OUT determines in which direction the
interface transmits data.
USER PORT (read operation)
The value is indicated in a window. A new
readout is started by pressing READ. Pressing
OK closes the window
00000000
USER PORT INPUT
READ
OK
IEC/IEEE-bus command: INP:UPOR:STAT ON
INP:UPOR?
USER PORT OUT (write operation)
The bit pattern selected with softkeys PORT 0 to
PORT 7 is output.
IEC/IEEE-bus command: OUTP:UPOR:STAT ON
PORT 0
0 1
The PORT 0 0/1 to PORT 7 0/1 softkeys determine the output
value for port 1 to 7. These softkeys are only available for
setting USER PORT OUT.
IEC/IEEE-bus command: OUTP:UPOR STAT ON
OUTP:UPOR #B10110010