R&S ESCI Receiver – SWEEP Key
1166.6004.12 4.45 E-2
Once measurement has been completed, the display of the measured values can be enlarged down to
the individual measured values by using the zoom function. Thus, each individual click interferer can be
evaluated in detail if necessary.
1PK
CLRWR
2QP
CLRWR
1s/
SGL
TRG
0
10
20
30
40
50
60
70
80
90
100
TRG 54.1 dBµV
RBW 120 kHz
MT 1 ms
PREAMP OFFAtt 10 dB
10 s
Freq 93 MHz
0 s
dBµV
Fig. 4-8 Scan in the time range
Time domain analysis is started with the RUN softkey. It can be performed as a single operation
(SINGLE) or as a repeating operation (CONTINUOUS). In SINGLE operation, time domain analysis
stops after the total run time has elapsed. In CONTINUOUS operation, time domain analysis can either
be interrupted with the HOLD SCAN softkey or stopped with the STOP SCAN softkey. The
measurement itself is continuous, i.e. if the end of the graph is reached and recording starts again at the
beginning of the graph, the measurement keeps running internally without any interruption.
Display of Measurement Results
Full-screen or split-screen mode can be selected for result display. In the full-screen mode, the result
diagram covers the whole screen area.
In split-screen mode, the upper half of the screen displays the frequency and level readout, i.e. a
bargraph. The diagram is displayed in the lower half
Up to 3 detectors can be measured simultaneously. They are assigned to traces 1 to 3. Since the
detectors are set only once, it is not possible to measure with different detectors in different the
subranges.