EasyManua.ls Logo

Rohde & Schwarz EMI Test ReceiverESCI - Power Measurements - Hardkey MEAS

Rohde & Schwarz EMI Test ReceiverESCI
495 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ESCI Analyzer - MEAS
1166.6004.12 4.145 E-1
Power Measurements – Hardkey MEAS
With its power measurement functions the R&S ESCI is able to measure all the necessary parameters
with high accuracy in a wide dynamic range.
A modulated carrier is almost always used (except e.g. SSB-AM) for high-frequency transmission of
information. Due to the information modulated upon the carrier, the latter covers a spectrum which is
defined by the modulation, the transmission data rate and the signal filtering. Within a transmission
band each carrier is assigned a channel taking into account these parameters. In order to ensure error-
free transmission, each transmitter must be conforming to the specified parameters. These include
among others:
the output power,
the occupied bandwidth, i.e. the bandwidth which must contain a defined percentage of the power
and
the power dissipation allowed in the adjacent channels.
Additionally the menu contains functions to determine the modulation depth of AM modulated signals
and to measure the 3
rd
order intercept point.
The measurements and the corresponding settings are selected in the MEAS menu.
MEAS menu:
SIGNAL
STATISTIC
MEAS
TRIG
TIME DOM
POWER
OCCUPIED
BANDWIDTH
SWEEP
BW
CHAN PWR
ACP
MODULATION
DEPTH
TOI
SELECT
MARKER
C/N
C/No
The MEAS key opens the menu to select and set the power
measurement.
The following measurements can be selected:
Power in the time domain (TIME DOM POWER)
Channel power and adjacent-channel power in the
frequency domain with a single carrier
(CHAN PWR ACP)
Channel power and adjacent-channel power in the
frequency domain with several carriers
(MULT CARR ACP)
Occupied bandwidth (OCCUPIED BANDWIDTH)
Carrier-to-noise ratio (C/N, C/No)
Amplitude probability distribution (SIGNAL STATISTICS)
Modulation depth (MODULATION DEPTH)
3
rd
order intercept (TOI)
The above measurements are carried out alternatively.

Table of Contents