R&S ESCI Configuration/Setup
1166.6004.12 4.221 E-3
ID STRING
USER
The ID STRING USER softkey opens an editor for entering a user-defined
response to the *IDN? query.
Max. length of output string: 36 characters
IEC/IEEE-bus command: --
GPIB
LANGUAGE
The GPIB LANGUAGE softkey opens a list of selectable remote-control
languages:
SCPI, 8560E, 8561E, 8562E, 8563E, 8564E, 8565E, 8566A, 8566B,
8568A, 8568B, 8594E
Note:
For 8566A/B, 8568A/B and 8594E, command sets A and B are available.
Command sets A and B differ in the rules regarding the command structure.
On switching between remote-control languages, the following settings or
changes will be made:
SCPI:
! The instrument will perform a PRESET.
8566A/B, 8568A/B, 8594E:
! The instrument will perform a PRESET.
! The following instrument settings will then be changed:
Model # of Trace
Points
Start Freq. Stop Freq. Ref Level Input
Coupling
8566A/B 1001 2 GHz 22 GHz 0 dBm DC (ESCI)
AC (ESPI)
8568A/B 1001 0 Hz 1.5 GHz 0 dBm AC
8560E 601 0 Hz 2.9 GHz 0 dBm AC
8561E 601 0 Hz 6.5 GHz 0 dBm AC
8562E 601 0 Hz 13.2 GHz 0 dBm AC
8563E 601 0 Hz 26.5 GHz 0 dBm AC
8564E 601 0 Hz 40 GHz 0 dBm AC
8565E 601 0 Hz 50 GHz 0 dBm AC
8594E 401 0 Hz 3 GHz 0 dBm AC
IEC/IEEE-bus command: SYST:LANG "SCPI" | "8560E" | "8561E" |
"8562E" | "8563E" | "8564E" |
"8565E" | "8566A" | "8566B" |
"8568A" | "8568B" | "8594E"