MEAS Key - Receiver R&S ESCI
1166.6004.12 4.30 E-2
FINAL MEAS menu (left side menu)
NO OF
PEAKS
The NO OF PEAKS softkey activates the entry field of the number of
subranges or peaks for the determination of the peak list. The range of
values is 1 to 500.
IEC/IEEE-bus command :CALC:PEAK:SUBR 1 to 500
PEAKS
SUBRANGES
The PEAKS SUBRANGES softkey defines the search method with which
maxima are searched for within a scan.
PEAKS Determination of a specific number of peak values
relative to the limit lines with the level values being
independent of their frequency spectral distribution.
SUBRANGES Generating subrange maxima, i.e. the interference
spectrum is further analyzed at frequencies with the
highest interference level of a frequency subrange.
IEC/IEEE-bus command :CALC:PEAK:METH SUBR|PEAK
MARGIN
The MARGIN softkey activates the entry field of the margin, i.e. of an
additional acceptance threshold for the determination of the peak list. The
limit line currently used is shifted by this amount for defining the maxima.
The range of values is -200 dB to 200 dB.
IEC/IEEE-bus command :CALC:PEAK:MARG -200dB to 200dB
FINAL
MEAS TIME
The FINAL MEAS TIME softkey activates the entry field of the time of final
measurement.
IEC/IEEE-bus command :SWEep:TIME:FME <num_value>
AUTOMATIC
FINAL
The AUTOMATIC FINAL softkey selects the automatic routine for the final
measurement. This routine is run according to the available frequency list
without user interaction..
IEC/IEEE-bus command :FME:AUTO ON