R&S ESCI Analyzer - Triggering the Sweep
1166.6004.12 4.99 E-1
Settings on R&S ESCI:
Conventions: [KEY] Menu called by this key. All information between the brackets refers to
this menu.
{Number} Numeric value to be entered for the parameter on hand.
SOFTKEY Softkey to be used for making a selection or entering a value.
[PRESET]
[FREQ: CENTER {802} MHz]
[SPAN {3.6} MHz]
[AMPT: REF LEVEL {0} dBm: RF ATTEN MANUAL: {10} dB]
[BW: RES BW MANUAL: {30} kHz]
[TRACE : TRACE 1 DETECTOR: RMS]
[SWEEP: SWEEPTIME MANUAL: {50} ms]
[TRIG: EXTERN
GATED TRIGGER;
GATE SETTINGS: GATE MODE EDGE; POLARITY POS
SWEEPTIME MANUAL {1} ms: GATE DELAY {300} µs:
GATE LENGTH: {250} µs]
The following figure shows the screen display for setting gate parameters. The vertical lines for gate
delay (GD) and gate length (GL) can be adjusted to the burst signal by entering numeric values or by
means of the rollkey.
Fig.4-12 Setting GATE DELAY and GATE LENGTH in time domain by means of lines GD and GL
On quitting the GATE SETTINGS menu, R&S ESCI returns to the previous screen.