R&S ESCI Analyzer - Traces
1166.6004.12 4.109 E-1
Example:
File contents Description
File header
Type;ESCI;
Version;1.00;
Date;01.Jul 1999;
Mode;Spectrum;
Center Freq;55000;Hz
Freq Offset;0;Hz
Span;90000;Hz
x-Axis;LIN;
Start;10000;Hz
Stop;100000;Hz
Ref.Level;-30;dBm
Level Offset;0;dB
Ref Position;75;%
y-Axis;LOG;
Level Range;100;dB
RF Att;20;dB
RBW;100000;Hz
VBW;30000;Hz
SWT;0.005;s
Trace Mode;AVERAGE;
Detector;SAMPLE;
Sweep Count;20;
Instrument model
Firmware version
Date of data set storage
Instrument mode
Center frequency
Frequency offset
Frequency range (0 Hz with zero span and statistics
measurements)
Scaling of x axis linear (LIN) or logarithmic (LOG)
Start/stop of the display range.
Unit:
Hz for span > 0, s for span = 0,
dBm/dB for statistics measurements
Reference level
Level offset
Position of reference level referred to diagram limits
(0% = lower edge)
Scaling of y axis linear (LIN) or logarithmic (LOG)
Display range in in y direction. Unit: dB with x axis LOG,
% with x axis LIN
Input attenuation
Resolution bandwidth
Video bandwidth
Sweep time
Display mode of trace:
CLR/WRITE,AVERAGE,MAXHOLD,MINHOLD
Detector set:
AUTOPEAK,MAXPEAK,MINPEAK,AVERAGE,
RMS,SAMPLE,QUASIPEAK
Number of sweeps set
Data section of the file
Trace 1:;;
x-Unit;Hz;
y-Unit;dBm;
Values; 501;
10000;-10.3;-15.7
10180;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Selected trace
Unit of x values:
Hz with span > 0; s with span = 0;
dBm/dB with statistics measurements
Unit of y values:
dB*/V/A/W depending on the selected unit with y axis
LOG or % with y axis LIN
Number of test points
Measured values:
<x value>, <y1>, <y2>
<y2> being available only with detector AUTOPEAK and
containing in this case the smallest of the two measured
values for a test point.