R&S ESCI Analyzer - Traces
1166.6004.12 4.113 E-1
TRACE-DETECTOR submenu
AUTO
SELECT
DETECTOR
SAMPLE
DETECTOR
AUTOPEAK
DETECTOR
MIN PEAK
DETECTOR
MAX PEAK
DETECTOR
RMS
DETECTOR
AVERAGE
DETECTOR
QPK
DETECTOR
The DETECTOR softkey opens a submenu for selecting the
detector for the selected trace. The softkey is highlighted if
the detector is not selected with
AUTO SELECT.
The detector can be selected independently for each trace.
The
AUTO SELECT mode selects the optimum detector for
each display mode of the trace (Clear/Write, Max Hold or Min
Hold).
The softkeys for the detectors are mutually exclusive
selection keys.
AUTO
SELECT
The AUTO SELECT softkey (= default setting) selects the
optimum detector for the set display mode of the trace
(Clear/Write, Max Hold and Min Hold) and the selected filter
mode (bandpass/FFT).
Trace display Detector (bandpass) Detector (FFT)
Clear/Write Auto Peak Max Peak
Average Sample Sample
Max Hold Max Peak Max Peak
Min Hold Min Peak Max Peak
The detector activated for the specific trace is identified in
the respective trace display field as follows:
Detector
Auto Peak AP
Max Peak PK
Min Peak MI
Average AV
RMS RM
Sample SA
Quasipeak QP
IEC/IEEE-bus command
DET:AUTO ON
DETECTOR
AUTO PEAK
The DETECTOR AUTOPEAK softkey activates the autopeak
detector.
IEC/IEEE-bus command
DET APE