R&S ESCI Analyzer - Traces
1166.6004.12 4.115 E-1
DETECTOR
AVERAGE
The DETECTOR AVERAGE softkey activates the average
detector.
In contrast to the rms detector, the average detector supplies
the linear average of all sampled level values during the
sweep of a pixel.
The same relations as for the rms detector apply (see
above).
IEC/IEEE-bus command
DET AVER
DETECTOR
QPK
The DETECTOR QPK softkey activates the quasipeak
detector.
This detector evaluates the sampled level values during the
sweep of a pixel like an analog voltmeter.
On switching the quasipeak detector on the video bandwidth
is automatically set to 10 MHz so as to exclude the influence
of the video filter on the signal evaluation.
IEC/IEEE-bus command
DET QPE