Calibration R&S ESCI
1166.6004.12 4.118 E-1
CAL ABORT
The CAL ABORT softkey interrupts the recording of correction data and
restores the last complete correction data set.
IEC/IEEE-bus command: CAL:ABOR
CAL CORR
ON OFF
The CAL CORR ON/OFF softkey switches the calibration data on/off.
ON The status message depends upon the results of the total
calibration.
OFF The message UNCAL appears in the status line.
IEC/IEEE-bus command: CAL:STAT ON
CAL
RESULTS
The CAL RESULTS softkey calls the CALIBRATION RESULTS table, which
shows the correction data found during calibration.
The CALIBRATION RESULTS table contains the following information:
date and time of last record of correction valuesKorrekturwertaufnahme
overall results of correction value record
list of found correction values according to function/module
The results have the following meaning:
PASSED calibration successful without any restrictions
CHECK deviation of correction value larger than expected, correction
could however be performed
FAILED deviations of correction value too large, no correction was
possible. The found correction data are not valid.
ABORTED calibration aborted
IEC/IEEE-bus command: CAL:RES?
PAGE UP
PAGE DOWN
The softkeys PAGE UP and PAGE DOWN scroll one page forward or
backward in the CALIBRATION RESULTS table. They have no function when
the table is closed.
IEC/IEEE-bus command: --