R&S ESCI MKR – Analyzer
1166.6004.12 4.121 E-1
Example:
[PRESET] R&S ESCI is set to the default setting.
[MKR] On calling the menu, MARKER 1 is switched on ('1' high-
lighted in the softkey) and positioned on the maximum
value of the trace. It is a normal marker and the
MARKER NORMAL softkey is highlighted.
[MARKER 2] MARKER 2 is switched on ('2' highlighted in the softkey).
It is automatically defined as a delta marker on
switching on so the DELTA is is
highlighted on softkey MARKER NORM DELTA. The fre-
quency and level of MARKER 2 with reference to
MARKER 1 are output in the marker info field.
[MARKER
NORM DELTA] The MARKER NORM DELTA softkey is highlighted.
MARKER 2 becomes a normal marker. The frequency
and level of MARKER 2 are output as absolute values in
the marker info field.
[MARKER 2] MARKER 2 is switched off. MARKER 1 is the active
marker for entry. The frequency and level of MARKER 1
are output in the marker info field.
IEC/IEEE-bus command: CALC:MARK ON;
CALC:MARK:X <value>;
CALC:MARK:Y?
CALC:DELT ON;
CALC:DELT:MODE ABS|REL
CALC:DELT:X <value>;
CALC:DELT:X:REL?
CALC:DELT:Y?
When several traces are being displayed, the marker is set to the maximum value (peak) of the active
trace which has the lowest number (1 to 3). In case a marker is already located there, it will be set to the
frequency of the next lowest level (next peak).
When the split-screen display mode is active, the marker will be placed in the active window. A marker
can only be enabled when at least one trace in the corresponding window is visible.
If a trace is turned off, the corresponding markers and marker functions are also deactivated. If the trace
is switched on again (VIEW, CLR/WRITE;..), the markers along with coupled functions will be restored
to their original positions provided the markers have not been used on another trace.