Limit Lines R&S ESCI
1166.6004.12 4.190 E-1
Unit - Select the vertical scale units for the limit line
The selection of units takes place in a selection box. The default setting is dBm.
UNITS
VERTICAL SCALE
dB
dBm
dBuV
dBmV
dBuA
dBpW
V
A
W
dBuV/MHz
dBmV/MHz
dBuA/MHz
%
IEC/IEEE-bus command: CALC:LIM3:UNIT DBM
Limit - Select upper/lower limit
A limit line can be defined as either an upper or lower limit.
IEC/IEEE-bus command: --
(defined by key words :UPPer or :LOWer)
Margin - Setting a margin.
The margin is defined as the signal-level distance to the limit line . When the
limit line is defined as an upper limit, the margin means that the level is below
the limit line. When the limit line is defined as a lower limit, the margin means
that the level is above the limit line. The default setting is 0 dB (i.e. no
margin).
IEC/IEEE-bus command: CALC:LIM3:UPP:MARG 10dB
CALC:LIM3:LOW:MARG 10dB