Limit Lines R&S ESCI
1166.6004.12 4.192 E-1
DELETE
VALUE
The DELETE VALUE softkey erases the data point (complete line) at the
cursor position. All succeeding data points are shifted down accordingly.
IEC/IEEE-bus command: --
SHIFT X
LIMIT LINE
The SHIFT X LIMIT LINE softkey calls an entry window where the complete
limit line may be shifted parallel in the horizontal direction.
The shift takes place according to the horizontal scale:
– in the frequency domain in Hz, kHz, MHz or GHz
– in the time domain in ns, µs, ms or s
In this manner, a new limit line can be easily generated based upon an
existing limit line which has been shifted horizontally and stored (SAVE LIMIT
LINE softkey) under a new name (NAME softkey).
IEC/IEEE-bus command: CALC:LIM3:CONT:SHIF 50KHz
SHIFT Y
LIMIT LINE
The SHIFT Y LIMIT LINE softkey calls an entry window where the complete
limit line may be shifted parallel in the vertical direction.
The shift takes place according to the vertical scale:
– for logarithmic units, relative, in dB
– for linear units, as a factor
In this manner, a new limit line can be easily generated based upon an
existing limit line which has been shifted vertically and stored (SAVE LIMIT
LINE softkey) under a new name (NAME softkey).
IEC/IEEE-bus command: CALC:LIM3:CONT:UPP:SHIF 20dB
CALC:LIM3:CONT:LOW:SHIF 20dB
SAVE
LIMIT LINE
The SAVE LIMIT LINE softkey stores the currently edited limit line . The
name can be entered in an input window (max. 8 characters)
IEC/IEEE-bus command: --