R&S ESCI Configuration/Setup
1166.6004.12 4.219 E-3
Factors - Selection of factors for the subrange
The permissible transducer factors for the marked subrange can be selected
in a selection box. Only factors matching with the unit of the set and fully
covering the selected subrange are permissible.
<NONE>
SELECT
Cable
Antenna1
After each change of range limits, R&S ESCI thus checks the factor list and,
if required, rebuilds it.
After reducing the start frequency or increasing the stop frequency of a range
it may happen that the factors defined for this range no longer fully cover the
range. These factors are deleted for this range when the transducer factor
table is opened next time.
A maximum of 4 transducer factors can be switched on at the same time in
each subrange. If none of them is switched on, 0 dB is assumed as a factor
for the whole subrange.
IEC/IEEE-bus command --
SAVE TRD
SET
The SAVE TRD SET softkey saves the changed table in a
file on the internal hard disk. If a transducer name with the
same name already exists, a corresponding query is
performed beforehand:
If the saved set is switched on, the new values will be used
immediately.
IEC/IEEE-bus command -- (executed automatically)