5.4 Programming the Transmitter
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
51
Select Pipe Class
1. Press the <Right Arrow> to select [Pick Pipe Class]. Press <Right Arrow> again and
scroll to desired Pipe Class.
2. Press <ENTER> to select.
3LSH'DWD
6LHPHQV &KDQQHO>@$%&
3LFN3LSH&ODVV
3L
FN3LSH&ODVV $6$&DUE6WHHO
6HOHFW3LSH6L]H &6
3LSH2'LQ!

3LSH0DWHULDO 6WHHO
:DOO7KLFNQHVV 
/LQHU0DWHULDO 1RQH
/LQHU7KLFNQHVV 
3. Pre-programmed Pipe Size and relevant pipe parameters will appear in menu cells. Press
<Right Arrow> and scroll to desired pipe size. Press <ENTER>. Enter dimensions
manually if pre-programmed dimensions do not match application.
Note
The DN sizes listed in the [Select Pipe Size] menu option list are referenced to DIN Table
2448. After selecting pipe size, check pipe OD and wall thickness for correct dimensions.
4. Press the <Left Arrow> and return to the main menu.
Select Liquid Class
1. Press the <Down Arrow> and scroll to [Application Data].
2. Press the <Right Arrow> to select [Liquid Class].
3. Press the <Right Arrow> again and scroll to desired liquid.
4. Press <ENTER> to save selection.
6HOHFW/LTXLG&ODVVIURP/LTXLG7DEOH
/LTXLG&ODVV :DWHU&)
8QL0DVV7DEOH 'LVDEOHG
7HPS5DQJH)WR)
3LSH&RQILJ)XOO\'HYHORSHG
$QRPDO\'LDPV 
$SSOLFDWLRQ'DWD
6LHPHQV&KDQQHO>@$%&
Select from list.