Page 12 of 16 7482255 Rev 01 Siemens
October 2002
Installing the G50 Third Array Port
5. Detach all transducers from the G50.
6. As depicted in Figure 5-11, loosen the two screws located at the bottom of
the front cover of the G50.
7. Pull the top of the front cover away from the G50 to release the three pairs
of magic fasteners (Figure 5-11).
8. Grasp the top of the front cover and one of the bumpers; pull and detach
the front cover from the G50.
9. Remove the two labels indicated in Figure 5-12 from the front of the probe
interface panel; these labels carry the message, “Do not open.”
10. Discard the two (“Do not open”) labels; they will not be reinstalled.
11. As indicated in Figure 5-12, remove the 12 screws (M4 X 8); retain these
screws for reinstallation during the reassembly of the G50.
12. After removing the 12 screws, pull the plastic loop (Figure 5-12) toward the
front of the G50 to remove the probe interface panel.
13. From the rear of the probe interface panel just removed, unscrew the two
screws to release the blank panel indicated in Figure 5-12.
Figure 5-11 Removing the Front Cover
Front cover
Magic fastener (3)
Screw (2)
Bumper
Main unit