Siemens 7482255 Rev 01 Page 9 of 42
October 2002
Chapter 8 - Software
18. Press the F4 key and select the Service menu item.
19. Enter the service password.
20. Select the DIMAQ-IP Maintenance button (Figure 8-10) located on the
Main Menu screen to display the Windows 2000 screen running on the
DIMAQ-IP module.
21. Enter the Windows 2000 Product Identification code and select the Ok but-
ton.
22. Wait until the DIMAQ-IP powers down, then de-energize the system.
23. Wait 15 seconds and re-energize the system.
24. Allow the DIMAQ-IP to boot completely; the boot process requires approx-
imately 5 minutes.
25. Press the Review button.
26. Observe the DIMAQ Study Screen with the HD radio button selected; this
indicates the success of the software update.
27. Create an image file, using any transducer, and assign a patient name and
patient identification number to it; this file will be used to verify the operation
of the CD.
28. Export the image file to the CD to verify its operation.
29. Verify that the exported image can be viewed by the system.
30. Delete the exported image from the HDD.
31. Reload the image file from the CD to the system.
32. Restore the Stress Echo feature key according to the instructions provided
in "Installing the Product Feature Keys" (page 8-10).
Note: On systems using the DIMAQ-IP software version 1.022, the first attempt
to write an image file to the CD will fail after new software has been installed. To
overcome this bug, attempt to export the image file to the CD; this attempt will fail.
Create a second image file and export both image files to a different CD: This
attempt to export the files to the CD should succeed. This bug was corrected by
version 1.023 and 1.2 DIMAQ-IP software.
!
Caution: Do not attempt to reload the DIMAQ-IP version 1.023 software to a
system that is operating with version 1.2 system software installed. This opera-
tion will fail and display the following message: “Please check the DIMAQ-IP
software version; DIMAQ- IP will shut down soon.”