4-45
Processing Instructions Operator Manual 920011-649
alternating with:
and printer records:
==========================
CYCLE PAUSE HH:MM:SS XM
==========================
2. If it is safe to resume cycle, display shows:
Press START touch pad to resume cycle. Printer records:
========================
CYCLE RESUME HH:MM:SS XM
========================
... and the cycle continues with the next phase in the sequence
until the cycle is completed.
...or...
3. Press STOP touch pad a second time to abort cycle.
Following Abort sequence, processor performs steps below one
or two times, depending on which cycle phase was in progress:
• chamber is filled with filtered hot tap water
• water is recirculated
• water is drained
NOTE: Procedure takes 7 to 15 minutes depending on cycle
phase in progress. Door remains locked until end of Abort
sequence(s).
4. If cycle cannot be resumed, display shows:
Press STOP touch pad to abort cycle.
During Abort sequence, display shows:
STOP PRESSED
CYCLE DELAYED
START TO RESUM
E
STOP TO CANCEL
STOP TO CANCEL
CANCELLING
CYCLE
WARNING – INEFFECTIVE
PROCESS HAZARD:
• Correct function of the
system is only ensured after
any diagnosed errors have
been found, corrected and a
successful processing cycle
has been completed.
• Endoscopes and accessories
are not adequately processed,
rinsed and air purged when
processing cycle is aborted.
REPROCESS LOAD.