EasyManua.ls Logo

SYSMEX XE-5000 - Page 153

SYSMEX XE-5000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-25
Revised April 2007 EU
8. Press the HELP key on the Main Unit panel keypad. The
HELP screen will appear.
9. If you select “OK” on the function menu of the HELP
screen, the Reagent setting screen will appear.
See “Chapter 9: 9.13 Replacing and registering reagents
for explanation of the Reagent setting screen.
STROMATOLYSER-NR(L) & STROMATOLYSER-NR(S)(SNR) replacement procedure
1. Prepare a new reagent and confirm that its expiration date
has not expired.
2. Remove the cap of a new STROMATOLYSER-NR(L).
3. Remove the cap of the empty STROMATOLYSER-NR(L).
Information
The FFS must be replaced every 2,000 times analysis,
an error will appear.
Always select “OK” after replacing the FFS.
SHUTDOWN HELP
SAMPLER
RETURN MORE
NUM./ALPH. ENTER
MANUAL
QZ
0 -/. C
PRS
1
TUV
2
WXY
3
GHI
4
JKL
5
MNO
6
7
ABC
8
DEF
9
Warning!
Before operating with the front cover open, be sure to
set the stop bar. Otherwise, the cover can drop down
and injure your head.
Caution!
• STROMATOLYSER-NR(L) and STROMATOLYSER-
NR(S) should be replaced at the same time.
Otherwise, correct analysis results may not be
obtained.
• If STROMATOLYSER-NR(S)(dye solution) spills,
wipe it off using cloth - preferably moistened with
alcohol. This is to prevent discoloration of the
instrument coat.

Table of Contents

Other manuals for SYSMEX XE-5000

Related product manuals