R3132 Series Spectrum Analyzer Operation Manual
3.3.18 POWER MEASURE Key (Power Measurement)
3-64
Spurious Time The Spuri Time menu and the spurious table information are dis-
played. (The spurious measurement table is not displayed when
there is no table data.)
When making spurious measurements while in zero span mode,
pass/fail judgment is repeated once for each frequency after the
setup conditions have been selected and a maximum of 10 signal
peaks have been searched for each frequency. When this function
is used with Average Power measurement, pass/fail judgments for
average powers are performed.
Table No. 1/2/3 Sets Table 1,2 or 3
1: Sets Table 1.
2: Sets Table 2.
3: Sets Table 3.
Edit Table Displays the Edit Table menu.
The editor of the selected table number is displayed. The frequen-
cies, RBW, sweep time and limit level which are used for pass/fail
judgments are entered. A maximum of 15 frequencies can be set.
Table No. 1/2/3 Sets Table 1,2 or 3.
1: Sets Table 1.
2: Sets Table 2.
3: Sets Table 3.
Insert Inserts the column at the cursor position.
Delete Deletes the column at the cursor position.
Table Init All data is cleared from the table.
Show Result Displays the Show Result menu and displays the measurement re-
sult in full screen mode. The setup conditions and detailed results
are displayed for each frequency.
Prev Result Displays the previous frequency result on the screen.
Next Result Displays the next frequency result on the screen.
Sweep Count ON/OFF Toggles the sweep count (used with the Max Hold mode, Min
Hold mode, Average function, Power Average function or Aver-
age Power measurement) on or off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3132 Series and is the answer not in the manual?

ADVANTEST R3132 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3132 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals