Maintenance and Troubleshooting 103
Troubleshooting the Protein Application
7
Symptoms (Protein)
Missing Peaks
Back to “Symptoms (Protein)” on page 89
Most probable causes Solution
Gel-dye mix was loaded in the destain
well.
Prepare a new chip.
Probable causes Solution
Fingerprint on focusing lens or on the
backside of the chip.
Clean lens as described in “Cleaning the Lens” on page 143.
Do not touch the underside of the chip.
Least probable causes Solution
Defective laser. Check the laser stability using the “Overview” on page 30. If the laser test fails,
contact Agilent Technologies at www.agilent.com/genomics/contact.
Autofocus failure. Check autofocus using the “Overview” on page 30. If autofocus test fails,
contact Agilent Technologies at www.agilent.com/genomics/contact.
High voltage power supply defective. Check high voltage stability using the “Overview” on page 30. If the high voltage
stability test fails, contact Agilent Te c hn o l ogi e s at
www.agilent.com/genomics/contact.

Table of Contents

Other manuals for Agilent Technologies 2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2100 and is the answer not in the manual?

Agilent Technologies 2100 Specifications

General IconGeneral
BrandAgilent Technologies
Model2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals