EasyManuals Logo

Agilent Technologies 2100 Maintenance And Troubleshooting Guide

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #105 background imageLoading...
Page #105 background image
Maintenance and Troubleshooting 105
Troubleshooting the Protein Application
7
Symptoms (Protein)
Poor Reproducibility
Back to “Symptoms (Protein)” on page 89
Most probable causes Solution
Wrong peak alignment. Check lower and upper marker assignment. Follow instructions for “Manual
Marker Assignment” on page 96.
Improper denaturation of sample(s). Use fresh sample aliquot. Heat samples with denaturing solution for 5 minutes at
100°C
Use 0.5 mL tubes for denaturing samples.
Samples not prepared similarly, i.e.
reducing agent (BME or DTT) was not
added to all samples.
Refer to the Kit Guide for proper sample reduction.
Dirty electrodes. Thoroughly clean the electrodes as described in “Maintenance of the Electrode
Cartridge” on page 115.
Probable causes Solution
Diluted samples are too old. Use diluted samples within one day.
Incompatible buffer component. See Protein Kit Guide for a list of compatible buffers and buffer compounds. For
an updated list, refer to the web site www.agilent.com/genomics/bioanalyzer.
If necessary, dilute, dialyze or desalt the sample.

Table of Contents

Other manuals for Agilent Technologies 2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2100 and is the answer not in the manual?

Agilent Technologies 2100 Specifications

General IconGeneral
BrandAgilent Technologies
Model2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals