Maintenance and Troubleshooting 113
Troubleshooting the Protein Application
7
Symptoms (High Sensitivity Protein)
Saturation of Lower Marker or Sample Peaks – Optical Signal too High
Back to “Symptoms (High Sensitivity Protein)” on page 112
Most probable causes Solution
Insufficient dilution of
ladder or samples.
Follow instructions in the Kit Guide. Dilution of the labeling reaction by 1:200 is recommended.
Probable causes Solution
Chip prepared with cold
reagents.
Prepare a new chip.
Allow all reagents and reagent mixes to warm up to room temperature for 30 minutes prior to use.
Store chips at room temperature.