EasyManuals Logo

Agilent Technologies 2100 Maintenance And Troubleshooting Guide

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
114 Maintenance and Troubleshooting
7
Troubleshooting the Protein Application
Symptoms (High Sensitivity Protein)
Low Signal Intensity
Back to “Symptoms (High Sensitivity Protein)” on page 112
Most probable causes Solution
Insufficient labeling of ladder or
samples.
Follow instructions in the Kit Guide.
Labeling occurs between pH 8-9.
Proteins must contain lysines for labeling reaction.
Check buffer compatibility in the “Compatibility List for the Labeling Reaction” in
the Kit Guide.
Insufficient sample present. Follow instructions for protein concentration in the Kit Guide.
Low abundant proteins may require pico labeling (Technical Note 5990-3703EN).

Table of Contents

Other manuals for Agilent Technologies 2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2100 and is the answer not in the manual?

Agilent Technologies 2100 Specifications

General IconGeneral
BrandAgilent Technologies
Model2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals