Maintenance and Troubleshooting 75
Troubleshooting the RNA Application
6
Symptoms (RNA)
Additional Saturating Bands
Back to “Symptoms (RNA)” on page 69
NOTE
Only present in RNA 6000 Pico or Small RNA assays.
Most probable causes Solution
Residual RNaseZAP on electrode pins. A dedicated electrode cassette for the use of RNA 6000 Pico and Small RNA
assays. This cassette should only be washed with water in the electrode cleaner
chip before and after each run. Only use RNaseZAP when decontaminating the
pins according to “How to Clean the Pin Set of the Electrode Cartridge” on
page 126.