EasyManua.ls Logo

Agilent Technologies 2100 - Page 76

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 Maintenance and Troubleshooting
6
Troubleshooting the RNA Application
Symptoms (RNA)
Degraded RNA Ladder and/or Samples
Back to “Symptoms (RNA)” on page 69
Most probable causes Solution
RNase contamination of the pin set. Decontaminate pin set. Follow decontamination procedure, see “How to Clean
the Pin Set of the Electrode Cartridge” on page 126.
Decontaminate pipettes and work space.
RNase contamination of chips and/or
reagents.
Prepare a new chip and fresh reagents.
Wear powder-free gloves when preparing the chip.
Decontaminate pipettes and work space.

Table of Contents

Other manuals for Agilent Technologies 2100

Related product manuals