84 Maintenance and Troubleshooting
6
Troubleshooting the RNA Application
Symptoms (RNA)
Back to “Symptoms (RNA)” on page 69
Least probable causes Solution
Changes in ambient temperature of
more than 5°C during the run.
Place the 2100 Bioanalyzer instrument in a thermally stable environment.
Autofocus failure or high voltage
power supply defective.
Check autofocus and high voltage power using the “Overview” on page 30. If a
diagnostic test fails, contact Agilent Tec h n ol o g ie s a t
www.agilent.com/genomics/contact.