Maintenance and Troubleshooting 95
Troubleshooting the Protein Application
7
Symptoms (Protein)
Unexpected Run Time
Unexpected Run Time
Back to “Symptoms (Protein)” on page 89
Most probable causes Solution
Lower and/or upper markers are called
incorrectly.
Turn off alignment and check which bands are the correct lower and upper
markers. For more details see “Manual Marker Assignment” on page 96 .
Least probable causes Solution
Bent electrode pin. Check if electrode pins are bent or damaged.
Replace electrophoresis cartridge.