Maintenance and Troubleshooting 97
Troubleshooting the Protein Application
7
Symptoms (Protein)
3 To adjust the lower marker, right click on the correct peak, and choose
Manually Set Lower Marker.
4 To adjust the upper marker, right click on the correct peak, and choose
Manually Set Upper Marker.
The data is properly aligned after the markers are called correctly.