27 MIL-STD-1553/ARINC 429 Triggering and Serial Decode
The base setting is used for data display in both the decode line and in
the Lister.
Label values are always displayed in octal, and SSM and SDI values are
always displayed in binary.
3 Press the Word Format softkey and specify the word decode format:
• Label/SDI/Data/SSM:
• Label - 8 bits.
• SDI - 2 bits.
• Data - 19 bits.
• SSM - 2 bits.
• Label/Data/SSM:
• Label - 8 bits.
• Data - 21 bits.
• SSM - 2 bits.
• Label/Data:
• Label - 8 bits.
• Data - 23 bits.
4 If the decode line does not appear on the display, press the [Serial] key
to turn it on.
5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
decode data.
You can use the horizontal Zoom window for easier navigation of the
decoded data.
See Also • "Interpreting ARINC 429 Decode" on page 419
• "ARINC 429 Totalizer" on page 420
• "Interpreting ARINC 429 Lister Data" on page 421
• "Searching for ARINC 429 Data in the Lister" on page 421
If the setup does not produce a stable trigger, the ARINC 429 signal may be slow enough
that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the
Mode softkey to set the trigger mode from Auto to Normal.