ATOMLAB 960 MEDICAL SPECTROMETER QUALITY ASSURANCE 5-19
12. Select <History> to display the “MDA History” screen.
13. Highlight an “MDA” and select <Reports> to go to the MDA Reports screen where you can
select: “Single Detail”, Summary History” or “Detail History”. The “Summary History” and
“Detail History” selections display the Date Range screen for the user to set the date range
to view and print.
14. Select <Back> to return to the Quality Assurance screen.
NOTE: You can print the MDA Report from the <Reports> function on the Home screen
.