91
Sp02 alarm information
When the alarm record function under certain menus is on, those physical alarms caused because
relevant parameters exceed the specified alarm limits will automatically output alarm parameter
values and relevant measurement waveforms.
Possible physical alarms, technical alarms and reminders during Sp02 module measurement are as
listed here:
Physical alarms:
Reminders Causes Alarming levels
SPO2 over-high SpO2 measurement value higher than the set
upper limit
User defined
SpO2 over-low SpO2 measurement value lower than the set
lower limit
User defined
PR over-high PR measurement value higher than the set
upper limit
User defined
PR over-low PR measurement value lower than the set
lower limit
User defined
Technical alarms
Reminders Causes Alar
m
Level
s
Remedial Measures
SPO2 Sensor
disconnected
SpO2 sensor drops down
from the patient or
monitor.
Low
Ensure the sensor is fixed on the
patient’s finger or other locus and
that the monitor is normally
connected with the cable.
SPO2 module
initialization wrong
SpO2 module wrong High
Stop use of SpO2 module
measurement function and advise
the biomedicine engineer or User
Service Dept.
SPO2 module
initialization wrong 1
SPO2 module
initialization wrong 2
SPO2 module
initialization wrong 3
SPO2 module
initialization wrong 4
SPO2 module
initialization wrong 5