Pet blood pressure cuff 4#
18 ~ 26 cm
10.6 cm 10.6 cm
Pet blood pressure cuff 5#
46 ~ 66 cm
21 cm 21 cm
14.2.2 Operational guide
1. Conduct one time of Automatic measurement
Enter the “NIBP Setting” menu, select a proper time interval at the “Time Interval” item, and press
the “START/STOP” button on the front panel. Then the system will start automatic pumping
measurement in the specified time interval.
Warning
If NIBP measurement under the Automatic mode lasts too long, body touching with the
cuff may have allergic purpuras, ischemia and neural injury. During monitoring on
patients, users should often check color, warmness and sensitiveness of remote body
parts. Once any abnormal phenomenon is found, users should put the cuff at another
location or immediately stop measuring blood pressure.
2. Stop automatic measurement
At any moment during the automatic measurement process, press the START/STOP button will
stop the automatic measurement.
3. Conduct one time of manual measurement
z Enter the “NIBP Setting” menu, select the “Time Interval” item and set its value
as ”Manual”, then press the START/STOP button on the front panel so as to start
manual measurement.
z During spare time of an automatic measurement, press the START/STOP button
will start a manual measurement; then if users press the START/STOP button again,
the manual measurement will stop and the automatic measurement will continue.
4. Conduct a manual measurement during automatic measurement process
Just press the START/STOP button on the control panel.
5. Stop a manual measurement
Re-press the START/STOP button on the control panel.
6. Conduct a continuous measurement
Enter the “NIBP Setting” menu and select the “Continuous” item to start a continuous
measurement, which will always last 10 minutes.