48
The patient’s vital signs are abnormal, relevant measures and treatment are immediately
required.
(3) Low level alarm :
The patient’s vital signs are abnormal,and relevant measures and treatment may be required.
All alarm levels for technical alarms and some physiological alarms have be set before the
monitors are delivered:, the users are not allowed to change them. but a certain degree of
physiological alarm can be modified.
8.3 Alarm Mode
When giving an alarm, the monitor will prompt the users in both sound and visual modes:
Lighting Alarm
Sound Alarm
Alarm
parameter’s flashing
In which light signal, sound signal and prompting message are differentiated with different alarm
levels.
8.4 Lighting Alarm
When giving an alarm, the alarm indicating lamp will show alarms in different levels with color
and flashing frequency.
High level alarm red with flash frequency of 2 per second.
Medium level alarm yellow with flash frequency of one every two seconds.
Low level alarm yellow without flash but with a continuous light.
8.5 Sound Alarm
The sound alarm is set to prompt the alarm in different levels of severity with different sounds.