99
Pressure change
Within certain time if the patient blood pressure immediately changes while users are
analyzing arterial pulse so as to obtain measurement values, measurement will be
unreliable or even impossible.
Serious shock
In case a patient is under serious shock or extreme low temperature, measurement will be
unreliable as reduction in blood flowing peripherally will result reduction in arterial
pulse.
HR limits
In case of HR lower than 40bpm or higher than 240bpm, no blood pressure measurement
can be done.
14.2.4 NIBP Parameter Setting & Adjustment
NIBP measurement results and relevant information are laid on screen as follows:
Fig 13-3
NIBP measurement results
14.3 NIBP setting in parameter area
Rotate the knob, move the cursor onto the NIBP hotkey in the parameter area, and then press the
knob to enter the NIBP Setting menu.
Measuring Mode
Measured Value
Measurement Time
BP Unit
Patient type
Prompt Area