sampling tube.
Warning
Do not use the ISA gas analyzer together with a quantitative spraying agent or
pulverization treatment; otherwise it may result in the clogging of the germ filter.
18.10 Discharging Waste Gases
When nitrous oxide and/or an anesthetic gas is used, you should prevent these gases from
polluting the operating room. Usually the gas discharging outlet should be connected to (via the
gas discharging pipe connected to the sample gas outlet of the host equipment):
A discharging system (used for discharging collected gases) or the patient circuit (used for the
back flowing of collected gases).
Warning
Anesthetics: When an anesthetic which is being used or a patient who recently used an
anesthetic is measured, the gas discharging hole on the module must be connected to a
waste gas processing system or the patient circuit (on the anesthesia machine or the
respirator), so as to prevent medical personnel from inhaling the anesthetic.
18.11 Consumables
The Nomoline sampling tube cannot be used repeatedly.
Replace the Nomoline sampling tube every two weeks or when “The sampling tube is clogged” is
displayed (based on whichever comes first).
18.12 Safety Symbol Information
Reminding
Symbol
Symbol Text, Color Code and
Description Text Format
Description
Warning: Supplementary text.
“Warning” refers to a dangerous situation which may
cause personal injury or death. A warning symbol
should be in conformity with: ISO 7010-W001.