98
Warning
If NIBP measurement under the Automatic mode lasts too long, body touching with the
cuff may have allergic purpuras, ischemia and neural injury. During monitoring on
patients, users should often check color, warmness and sensitiveness of remote body
parts. Once any abnormal phenomenon is found, users should put the cuff at another
location or immediately stop measuring blood pressure.
7. Stop continuous measurement
At any moment during the continuous measurement process, press the START/STOP button will
stop the continuous measurement.
Caution
In case of suspecting reading accuracy, users should take possible methods to check life
signs of patients before checking the monitor,
Warning
In case any liquid is sprayed onto the equipment or its accessories, especially when the
liquid may enter the tube or monitor, please contact with the maintenance department
of your hospital.
14.2.3 Measurement restriction
Vibration measurement has its restriction subject with patient conditions. This measurement
method looks for regular pulse waves generated from arterial pressure, so when patient conditions
make this wave detection method hard to work, measured values are no more reliable and
measurement time last longer. Users must understand the following cases will disturb the
measurement method, making measured press unreliable or measurement time extended. In such
cases, patient conditions disable measurement to be continued.
Patient movement
In case a patient is moving, shaking or convulsing, measurement will be unreliable or
even impossible; as such scenarios will disturb detection of arterial pulse and extend
measurement time.
Arrhythmia
In case a patient shows irregular heartbeats resulted from arrhythmia, measurement will
be unreliable or even impossible, while measurement time will also be extended.
Heart-lung machine
If a patient is connected with an artificial heart-lung machine, measurement can’t be
realized.