8.9 Setting the upper and lower limits of alarm
Take ECG for example:
a) Select ECG in the parameter area;
b) In the popped-out menu ECG Setting, select the HR ALM SETUP;
c) Set the upper and lower limits of HIGH LIMIT as 120bpm and 50bpm respectively;
d) Set the upper and lower limits of MID LIMIT as 110bpm and 60bpm respectively
e) Set the upper and lower limits of LOW LIMIT as 100bpm and 70bpm respectively
f) Upon completion of setting, press the key ā ā to exit.
Warning
When setting the upper and lower limits of alarm, decide whether the patient you are
monitoring is an big or small animal, and the limit should be set according to clinical
requirements.If the set limit value exceeds the alarm limit, it will easily lead to alarm
system failure.