140
17.7 Alarm information and prompt information
17.7.1 Alarm information
When alarm record switches in related menus are turned on, physiological alarms given when
parameters exceeds alarm limits will trigger the recorder to automatically output alarmed
parameters and related measured waveforms.
Possible physiological alarms, technical alarms and prompt messages in IBP module measurement
are listed in the following tables:
Physiological alarms:
PROMPT
MESSAGE
CAUSE ALARM
LEVEL
IS too high Measured SP value is higher than the set upper
alarm limit.
User Optional
IS too low Measured SP value is lower than the set lower
alarm limit.
User Optional
ID too high Measured DP value is higher than the set upper
alarm limit.
User Optional
ID too low Measured DP value is lower than the set lower
alarm limit.
User Optional
IM too high Measured MP value is higher than the set upper
alarm limit.
User Optional
IM too low Measured MP value is lower than the set lower
alarm limit.
User Optional
Technical alarms:
PROMPT
MESSAGE
CAUSE ALARM
LEVEL
SOLUTION
IBP lead detached IBP cable is
detached from the
monitor
Low Make sure the cable is connected
reliably.
IBP module
initialization
wrong
IBP measurement
module has faults
High Suspend the IBP measurement function
and inform biomedical engineers or
maintenance workers.
17.8 Maintenance and Cleaning