with the sampling pipe.
Attention
If the interface shows the prompt of ‘no connection to the oxygen sensor’, please reinstall the
oxygen sensor.
Attention
The end of the gas circuit adapter which connects the gas sampling pipe should point
upward so as to prevent the condensing water drops entering the gas sampling pipe and
blocking it up.
Warning
z Hang the external CO
2
analyzer onto the CO
2
bracket on the rear casing of the
instrument; prevent the dropping damage of the CO
2
module.
z Unless HME is used to protect the IRMA probe, the state indicating LED should face
upward all the time during IRMA probe placement.
z Do not pull the cable of ISA By-flow Gas Analyzer.
z Do not operate the ISA By-flow Gas Analyzer in the environment beyond the designated
working temperature.
z Make sure all connections are firm and reliable. Any leakage will result in the inclusion
of ambient air in the patients respiratory gas, which leads to a wrong reading.
16.5 CO2 Settings in Parameter Area
Rotate the rotary shuttle button to move the cursor on the display interface to the CO2 hotkey in
the parameter area, and then press the rotary shuttle button to enter the menu “CO2 Settings”.