104
In about 20 seconds, the system will automatically open the air bleeder, indicating completion of
gas leakage measurement.
No reminder information displaying on the NIBP parameter area doesn’t mean no gas leakage
within the system. Display of “Pump leaking…” means there is possible gas leakage with the gas
route, in such case operators should check whether there is any loose connection, and after making
sure of no more loose connection re-do the gas leakage detection; if the error reminder is still
displayed, please contact with the manufacturer for repair
Fig 13-7 Connection for NIBP Gas Leakage Detection
14.7 NIBPAlarms & Reminders
Provided that the alarming record function under relevant menu is turned on, those physical
alarms activated because parameters exceed alarming limits may activate the recorder to
automatically output alarming parameter values and the relevant measurement waveforms.The
following table lists various possible alarms during NIBP measurement process.
Physical alarms:
Reminders Causes Alarming levels
NS over-high Measured NIBP systolic pressure higher than specified
alarming upper limit
User defined
NS over-low Measured NIBP systolic pressure lower than specified
alarming lower limit
User defined
ND over-high Measured diastolic pressure higher than specified alarming
upper limit
User defined
ND over-low Measured diastolic pressure lower than specified alarming
lower limit
User defined
NM over-high Measured average pressure higher than specified alarming
upper limit
User defined
NM over-low Measured average pressure lower than specified alarming
lower limit
User defined
Technical alarms 1 (displayed in the monitor information area):
Reminders Causes Alarming Solution
Gas pipe
Cylinder
Cuff