92
SPO2 module
initialization wrong 6
SPO2 module
initialization wrong 7
SPO2 module
initialization wrong 8
SPO2 module
communication stop
SpO2 module wrong or
communication mistaken
High
Stop use of SpO2 module
measurement function and advise
the biomedicine engineer or User
Service Dept.
SPO2 module
communication wrong
SPO2 alarm limit wrong Function/safety faulty High
Stop use of SpO2 module
measurement function and advise
the biomedicine engineer or User
Service Dept.
PR alarm limit wrong Function/safety faulty High
Stop use of SpO2 module
measurement function and advise
the biomedicine engineer or User
Service Dept.
Prompt Message (including general warnings):
Reminders Causes Alarm Levels
SPO2 Measurement out of
scope
SpO2 measured value is out of range High
PR Measurement out of
scope
PR measured value is out of range High
Seek pulses SpO2 module seeks pulse. No alarm
Pulse not found SpO2 module can’t find SpO2 signal. High
13.9 Maintenance & Cleaning
Warning
Users must turn off the equipment and shut down the AC power supply before cleaning
the monitor or the connected sensor.
Caution
Please don’t sterilize the senor with pressure.
Please don’t soak the sensor in the liquid.