137
AGJUST
z DEFAULT: checked into the IBP <1,2> Default Settings dialog box, select "No" to
abandon the current operation, the system remains the original configuration
unchanged, select "Yes", and will be using the default settings,the original
configuration will be overwritten.
IBP PRESSURE ZERO
z CH1 ZERO: the IBP1 and IBP2 module into the socket, the need for invasive
pressure zeroing.
z CH2 ZERO: the IBP1 and IBP2 module into the socket, the need for invasive
pressure zeroing.
CAL(calibration)
z CH1 CAL VALUE: Adjustable from 80 to 300 values.
z CH2 CAL VALUE: Adjustable from 80 to 300 values.
z CH1 ADJUST
z CH2 ADJUST
Warning
When setting alarm limits, users should confirm the item to be set.
17.5 Zero calibration of sensor:
Press the “IBP zero calibration” key with the rotary knob and the system will begin zero
calibration.
Zero calibration consideration:
Before zero calibration, close the three-way stop cock at the patient side.
Before zero calibration, the sensor should communicate with the atmosphere.
The sensor must be positioned at the same level with the heart, approximately on the
midaxillary line.
Zero calibration should be done before monitoring start and at lease once per day (zero
calibration must be done each time after the cable is plugged or pulled.)