EasyManua.ls Logo

Comen C80 - Ibp<1,2>Setup in the Measurement Setup

Comen C80
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
136
z
SWEEP:12.5mm/s,25mm/s
z
FILTER: smooth, normal ,no filter
z
WAVE COLOR:green ,cyan , red ,yellow ,white ,blue , purple
z
WAVE TYPE:line,fill
17.4.3 IBP<1,2>setup in the measurement setup
Select the “Module Setup” option in the “monitor setup” menu and set the CO2 on-off to be on.
The following figure (the present figure is in demonstrating mode) will come out:
Press the knob to enter the “IBP selection” menu.
Fig 17-4 IBP Parameter Setting Menu
Settings can be done on the following items:
IBP<1,2>SETUP:
z ALM ON/OFF: see this chapter parameter area IBP <1,2> setup the ALM ON/OFF.
z ALM LEV: see this chapter parameter area IBP <1,2> setup the ALM LEV
z ALM REC: see this chapter parameter area IBP <1,2> setup the ALM REC
z SWEEP: see this chapter waveform area IBP <1,2> setup the SWEEP
z WAVE COLOR: see this chapter waveform area IBP <1,2> setup the WAVE
COLOR.
z UNIT : mmHg,kPa,cmH2O
z FILTER: see this chapter waveform area IBP <1,2> setup the FILTER.
z CH Press Setup: see this chapter waceform area IBP <1,2> setup the CH Press
Setup
z ALARM SETUP: see this chapter parameter area IBP <1,2> setup the ALARM
SETUP
z SCALE AGJUST see this chapter parameter area IBP <1,2> setup the SCALE

Table of Contents