166
Appendix II. Specification
Monitor type
Name Type
classified by electric shock
protection
belongs to Class I, Type CF application part, Type BF application
part anti-defibrillator devices with internal power supply Among
which except for the cardiac electrophysiologic detecting part, which
belongs to Type CF application part, other detecting part belong to
Type BF application part
Safety standards
IEC 60601-1; IEC 60601-2-27 and GB 9706.25;
IEC 60601-1-2 and YY 0505; IEC 60601-2-30 and YY 0667;
IEC 60601-2-49 and YY 0668; YY 91079.
anti-leakage extent Common seal apparatus, without anti-leakage character
Disinfection / sterilization
methods
Refer to Chapter 10 -16for details
work mode Continuous
2. Monitor Specifications
(1) Size and weight
Name Specifications
Size and weight
Size: 344.5mm×291 mm×165mm
Weight: 4. 5kg
(2) Environment requirement
Name Specifications
Working
environment
Environment temperature range 5℃~40℃
Relative humidity range ≤80%
Atmospheric pressure range 86kPa~106kPa
Power frequency 50Hz/60Hz
Input 70VA
Transport: Must avoid severe shock ,vibration, rain and snow during transport
Storage
Packed monitors must be stored in well ventilated rooms with -10℃~+40℃
temperature, relative humidity no more than 80%, and without corrosive gases
(3) Displayer specification
Name Specifications
display 12.1 inches color LCD displayer