Warning
If bubbles are found in the pressure pipe or the sensor, flush the system with the
perfusion liquid.
1. Position the sensor on the same level as the heart, approximately on the
midaxiallary line
2. Confirm correct ruler names have been chosen. See the following section for
details.
3. Do zero-adjustment of the sensor. See the following section for details.
Fig 16-1 IBP Monitoring
17.4 IBP Menu
17.4.1 IBP<1,2>setup in the parameter area
Rotate the knob and move the cursor to the IBP hotkey in the parameter area of the screen.